 +==== Detector ====
 +
 +Open Firefox and connect to 192.168.127.50
 +
 +{{:crystallography:collection:frx:framegrabber_1.png?600|}}
 +
 +Select 'system settings'. Username = eiger, PW = #EIGER_Detector#
 +
 +In the lefthand column select 'EIGER' and then 'DCU-Control'
 +
 +{{:crystallography:collection:frx:framegrabber_2.png?600|}}
 +
 +select "Power OFF" to shut down the framegrabber.
 +
 +{{:crystallography:collection:frx:framegrabber_3.png?600|}}
 +
 +Switch OFF the power supply for the EIGER.
 +
 +{{:crystallography:collection:frx:eiger_power_supply.jpg?600|}}
 +
 +Switch OFF the EIGER water chiller.
 +
 +{{:crystallography:collection:frx:eiger_chiller.jpg?400|}}
  
 ---- ----
Line 5: Line 31:
 ==== FR-X ==== ==== FR-X ====
  
-Decrease gradually the the current 66 to 10 mA and then voltage from 45 kV to 20 kV. First step 66->60 mA, (45->40 kV). Wait 10 s. Then decrease by steps of 10. (Lower the current first)+Decrease gradually the the current 66 to 10 mA and then voltage from 45 kV to 20 kV. First step 66->60 mA, (45->40 kV). Wait 10 s. Then decrease by steps of 10. (Lower the current first.) 
 + 
 +{{:crystallography:collection:frx:xg.png?600|}}
  
 Turn X-RAY OFF on the XG Control panel. Wait 1h. Turn X-RAY OFF on the XG Control panel. Wait 1h.
Line 12: Line 40:
  
 POWER OFF on the FR-X main panel. POWER OFF on the FR-X main panel.
 +
 +{{:crystallography:collection:frx:frx_front.jpg?600|}}
  
 Switch OFF at the back of the FR-X. Switch OFF at the back of the FR-X.
 +
 +{{:crystallography:collection:frx:frx_back.jpg?400|}}
  
 Switch the main power source OFF. Switch the main power source OFF.
 +
 +{{:crystallography:collection:frx:main_power_supply.jpg?600|}}
 +
 +Switch OFF AFC-11 Goniometer controller.
 +
 +{{:crystallography:collection:frx:afc11_closeup.jpg?600|}}
 +
 +Switch OFF the two fans of AFC-11.
 +
 +{{:crystallography:collection:frx:acf11_back.jpg?400|}}
 +
 +Switch OFF the vacuum pump below the Cobra Controller.
 +
 +{{:crystallography:collection:frx:pump1.jpg?250|}}
 +{{:crystallography:collection:frx:pump1_closeup.jpg?250|}}
 +
 +Switch OFF the vacuum pump below the BioSAXS.
 +
 +{{:crystallography:collection:frx:pump2.jpg?250|}}
 +{{:crystallography:collection:frx:pump2_closeup.jpg?350|}}
  
 ---- ----
Line 21: Line 73:
 ==== Cryostream ==== ==== Cryostream ====
  
-Program / Ramp (change with dial) / Enter+On the Cobra Controller :
  
-Choose the speed (up to 360K/h) / Enter +{{:crystallography:collection:frx:cobra_front.jpg?600|}}
-Target / Ambient / Enter +
-Enter / Enter +
-Start+
  
-When ambient temperature is reached, stop +  * Program / Ramp (change with dial) / Enter 
-Turn OFF at the back+  * Choose the speed (up to 360K/h) / Enter 
 +  * Target / 293 K (Ambient) / Enter 
 +  * Start 
 +  * When ambient temperature is reached, press Stop 
 + 
 +Switch OFF at the back
 + 
 +{{:crystallography:collection:frx:cobra_back.jpg?600|}} 
 + 
 +Switch OFF the helium compressor. 
 + 
 +{{:crystallography:collection:frx:helium.jpg?300|}} 
 +{{:crystallography:collection:frx:helium_back.jpg?250|}}
  
 ---- ----
  
-==== Engine Room ====+==== Technical Room E0044 ==== 
 + 
 +{{:crystallography:collection:frx:technical_room.jpg?400|}} 
 + 
 +Switch OFF the nitrogen compressor using the touch panel. 
 +Switch OFF the compressor on the back of the device. 
 + 
 +{{:crystallography:collection:frx:n2.jpg?150|}} 
 +{{:crystallography:collection:frx:n2_closeup.jpg?400|}} 
 + 
 +Switch OFF the main switch of the water cooling device (Haskins). 
 + 
 +{{:crystallography:collection:frx:haskins.jpg?200|}} 
 +{{:crystallography:collection:frx:haskins_closeup.jpg?400|}} 
 + 
 +Close the water supply. (red valve) 
 + 
 +{{:crystallography:collection:frx:red_valve.jpg?400|}} 
 + 
 +---- 
 + 
 +==== EIGER Computer ==== 
 + 
 +Shut down the computer. 
 + 
 +{{:crystallography:collection:frx:pc.jpg?600|}} 
 + 
 +---- 
 + 
 +==== BioSAXS ==== 
 + 
 +Switch OFF 
 + 
 +{{:crystallography:collection:frx:saxs_control_1.jpg?600|}} 
 + 
 +Switch OFF at the back. 
 + 
 +{{:crystallography:collection:frx:saxs_control_2.jpg?600|}} 
 + 
 +Remove the small middle screw to access the Pilatus 100K detector and switch OFF. 
 + 
 +{{:crystallography:collection:frx:pilatus100k_cover.jpg?200|}} 
 +{{:crystallography:collection:frx:pilatus100k.jpg?400|}} 
 + 
 +---- 
 + 
 +==== BioSAXS Computers ==== 
 + 
 +Shut down the SAXS PC. 
 + 
 +{{:crystallography:collection:frx:saxs_pc.jpg?600|}} 
 + 
 +Switch the display and control to the framegrabber with the Switch. 
 + 
 +{{:crystallography:collection:frx:saxs_pc_switch.jpg?400|}} 
 + 
 +Shut down the SAXS framegrabber.
  
-Switch OFF the nitrogen compressor in the engine room using the touch panel. 
  
-Switch OFF the main switch of the water cooling device in the engine room.  