User Tools

Site Tools


crystallography:collection:frx:off

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
crystallography:collection:frx:off [2017/12/19 14:14] bonnefoncrystallography:collection:frx:off [2023/11/01 20:19] (current) – external edit 127.0.0.1
Line 5: Line 5:
 ==== Detector ==== ==== Detector ====
  
-Open firefox and connect to 192.168.127.50+Open Firefox and connect to 192.168.127.50
  
-Select system settings. Username = eiger, PW = #EIGER_Detector#+{{:crystallography:collection:frx:framegrabber_1.png?600|}} 
 + 
 +Select 'system settings'. Username = eiger, PW = #EIGER_Detector#
  
 In the lefthand column select 'EIGER' and then 'DCU-Control' In the lefthand column select 'EIGER' and then 'DCU-Control'
 +
 +{{:crystallography:collection:frx:framegrabber_2.png?600|}}
  
 select "Power OFF" to shut down the framegrabber. select "Power OFF" to shut down the framegrabber.
 +
 +{{:crystallography:collection:frx:framegrabber_3.png?600|}}
  
 Switch OFF the power supply for the EIGER. Switch OFF the power supply for the EIGER.
Line 25: Line 31:
 ==== FR-X ==== ==== FR-X ====
  
-Decrease gradually the the current 66 to 10 mA and then voltage from 45 kV to 20 kV. First step 66->60 mA, (45->40 kV). Wait 10 s. Then decrease by steps of 10. (Lower the current first)+Decrease gradually the the current 66 to 10 mA and then voltage from 45 kV to 20 kV. First step 66->60 mA, (45->40 kV). Wait 10 s. Then decrease by steps of 10. (Lower the current first.)
  
-{{:crystallography:collection:frx:control_1.jpg?600|}}+{{:crystallography:collection:frx:xg.png?600|}}
  
 Turn X-RAY OFF on the XG Control panel. Wait 1h. Turn X-RAY OFF on the XG Control panel. Wait 1h.
Line 51: Line 57:
 Switch OFF the two fans of AFC-11. Switch OFF the two fans of AFC-11.
  
 +{{:crystallography:collection:frx:acf11_back.jpg?400|}}
  
- +Switch OFF the vacuum pump below the Cobra Controller.
-Switch OFF vacuum pump below Cobra.+
  
 {{:crystallography:collection:frx:pump1.jpg?250|}} {{:crystallography:collection:frx:pump1.jpg?250|}}
 {{:crystallography:collection:frx:pump1_closeup.jpg?250|}} {{:crystallography:collection:frx:pump1_closeup.jpg?250|}}
  
-Switch OFF vacuum pump below BioSAXS+Switch OFF the vacuum pump below the BioSAXS.
  
 {{:crystallography:collection:frx:pump2.jpg?250|}} {{:crystallography:collection:frx:pump2.jpg?250|}}
Line 66: Line 72:
  
 ==== Cryostream ==== ==== Cryostream ====
 +
 +On the Cobra Controller :
  
 {{:crystallography:collection:frx:cobra_front.jpg?600|}} {{:crystallography:collection:frx:cobra_front.jpg?600|}}
  
-Program / Ramp (change with dial) / Enter+  * Program / Ramp (change with dial) / Enter 
 +  * Choose the speed (up to 360K/h) / Enter 
 +  * Target / 293 K (Ambient) / Enter 
 +  * Start 
 +  * When ambient temperature is reached, press Stop
  
-Choose the speed (up to 360K/h) / Enter +Switch OFF at the back.
- +
-Target / Ambient / Enter +
- +
-Enter / Enter +
- +
-Start +
- +
-When ambient temperature is reached, press Stop +
- +
-Switch OFF at the back+
  
 {{:crystallography:collection:frx:cobra_back.jpg?600|}} {{:crystallography:collection:frx:cobra_back.jpg?600|}}
  
-Switch off the helium compressor+Switch OFF the helium compressor.
  
 {{:crystallography:collection:frx:helium.jpg?300|}} {{:crystallography:collection:frx:helium.jpg?300|}}
Line 92: Line 94:
 ---- ----
  
-==== Technical Room ====+==== Technical Room E0044 ====
  
 {{:crystallography:collection:frx:technical_room.jpg?400|}} {{:crystallography:collection:frx:technical_room.jpg?400|}}
  
-Switch OFF the nitrogen compressor in the engine room using the touch panel.+Switch OFF the nitrogen compressor using the touch panel
 +Switch OFF the compressor on the back of the device.
  
 {{:crystallography:collection:frx:n2.jpg?150|}} {{:crystallography:collection:frx:n2.jpg?150|}}
 {{:crystallography:collection:frx:n2_closeup.jpg?400|}} {{:crystallography:collection:frx:n2_closeup.jpg?400|}}
  
-Switch OFF the main switch of the water cooling device (Haskins) in the technical room.+Switch OFF the main switch of the water cooling device (Haskins).
  
 {{:crystallography:collection:frx:haskins.jpg?200|}} {{:crystallography:collection:frx:haskins.jpg?200|}}
Line 112: Line 115:
 ---- ----
  
-==== Computer ====+==== EIGER Computer ====
  
 Shut down the computer. Shut down the computer.
  
 {{:crystallography:collection:frx:pc.jpg?600|}} {{:crystallography:collection:frx:pc.jpg?600|}}
 +
 +----
 +
 +==== BioSAXS ====
 +
 +Switch OFF
 +
 +{{:crystallography:collection:frx:saxs_control_1.jpg?600|}}
 +
 +Switch OFF at the back.
 +
 +{{:crystallography:collection:frx:saxs_control_2.jpg?600|}}
 +
 +Remove the small middle screw to access the Pilatus 100K detector and switch OFF.
 +
 +{{:crystallography:collection:frx:pilatus100k_cover.jpg?200|}}
 +{{:crystallography:collection:frx:pilatus100k.jpg?400|}}
 +
 +----
 +
 +==== BioSAXS Computers ====
 +
 +Shut down the SAXS PC.
 +
 +{{:crystallography:collection:frx:saxs_pc.jpg?600|}}
 +
 +Switch the display and control to the framegrabber with the Switch.
 +
 +{{:crystallography:collection:frx:saxs_pc_switch.jpg?400|}}
 +
 +Shut down the SAXS framegrabber.
 +
 +
crystallography/collection/frx/off.1513692841.txt.gz · Last modified: (external edit)