Line 33: Line 33:
 Decrease gradually the the current 66 to 10 mA and then voltage from 45 kV to 20 kV. First step 66->60 mA, (45->40 kV). Wait 10 s. Then decrease by steps of 10. (Lower the current first.) Decrease gradually the the current 66 to 10 mA and then voltage from 45 kV to 20 kV. First step 66->60 mA, (45->40 kV). Wait 10 s. Then decrease by steps of 10. (Lower the current first.)
  
-{{:crystallography:collection:frx:control_1.jpg?600|}}+{{:crystallography:collection:frx:xg.png?600|}}
  
 Turn X-RAY OFF on the XG Control panel. Wait 1h. Turn X-RAY OFF on the XG Control panel. Wait 1h.
Line 57: Line 57:
 Switch OFF the two fans of AFC-11. Switch OFF the two fans of AFC-11.
  
 +{{:crystallography:collection:frx:acf11_back.jpg?400|}}
  
 Switch OFF the vacuum pump below the Cobra Controller. Switch OFF the vacuum pump below the Cobra Controller.
Line 79: Line 79:
   * Program / Ramp (change with dial) / Enter   * Program / Ramp (change with dial) / Enter
   * Choose the speed (up to 360K/h) / Enter   * Choose the speed (up to 360K/h) / Enter
-  * Target / Ambient / Enter +  * Target / 293 K (Ambient/ Enter
-  * Enter / Enter+
   * Start   * Start
   * When ambient temperature is reached, press Stop   * When ambient temperature is reached, press Stop
Line 88: Line 87:
 {{:crystallography:collection:frx:cobra_back.jpg?600|}} {{:crystallography:collection:frx:cobra_back.jpg?600|}}
  
-Switch off the helium compressor.+Switch OFF the helium compressor.
  
 {{:crystallography:collection:frx:helium.jpg?300|}} {{:crystallography:collection:frx:helium.jpg?300|}}
Line 95: Line 94:
 ---- ----
  
-==== Technical Room ====+==== Technical Room E0044 ====
  
 {{:crystallography:collection:frx:technical_room.jpg?400|}} {{:crystallography:collection:frx:technical_room.jpg?400|}}
  
 Switch OFF the nitrogen compressor using the touch panel. Switch OFF the nitrogen compressor using the touch panel.
 +Switch OFF the compressor on the back of the device.
  
 {{:crystallography:collection:frx:n2.jpg?150|}} {{:crystallography:collection:frx:n2.jpg?150|}}
Line 125: Line 125:
 ==== BioSAXS ==== ==== BioSAXS ====
  
 +Switch OFF
 +
 +{{:crystallography:collection:frx:saxs_control_1.jpg?600|}}
 +
 +Switch OFF at the back.
 +
 +{{:crystallography:collection:frx:saxs_control_2.jpg?600|}}
 +
 +Remove the small middle screw to access the Pilatus 100K detector and switch OFF.
 +
 +{{:crystallography:collection:frx:pilatus100k_cover.jpg?200|}}
 +{{:crystallography:collection:frx:pilatus100k.jpg?400|}}
  
 ---- ----
  
-==== BioSAXS Computer ====+==== BioSAXS Computers ==== 
 + 
 +Shut down the SAXS PC. 
 + 
 +{{:crystallography:collection:frx:saxs_pc.jpg?600|}} 
 + 
 +Switch the display and control to the framegrabber with the Switch. 
 + 
 +{{:crystallography:collection:frx:saxs_pc_switch.jpg?400|}} 
 + 
 +Shut down the SAXS framegrabber. 
  
crystallography/collection/frx/off.1513696457.txt.gz · Last modified: (external edit)