crystallography:collection:frx:on
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| crystallography:collection:frx:on [2018/01/04 08:25] – bonnefon | crystallography:collection:frx:on [2023/11/01 20:19] (current) – external edit 127.0.0.1 | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| Turn ON FR-X | Turn ON FR-X | ||
| - | ===== Turn in-house FR-X OFF ===== | + | ===== Turn in-house FR-X ON ===== |
| ---- | ---- | ||
| Line 8: | Line 8: | ||
| Turn the tap on the main water inlet to the ON position. | Turn the tap on the main water inlet to the ON position. | ||
| + | |||
| + | {{: | ||
| Switch ON the water chiller (HASKRIS) using the main switch only. | Switch ON the water chiller (HASKRIS) using the main switch only. | ||
| Wait until the temperature is stabilized at 17°C. | Wait until the temperature is stabilized at 17°C. | ||
| + | |||
| + | {{: | ||
| + | {{: | ||
| Switch ON the nitrogen compressor at the back, then start the nitrogen flow with the touch panel. | Switch ON the nitrogen compressor at the back, then start the nitrogen flow with the touch panel. | ||
| After a 3 min idle wait, the pump should stabilize in 30 min. | After a 3 min idle wait, the pump should stabilize in 30 min. | ||
| - | Wait for 1 h. | + | Wait for 1 h. |
| + | |||
| + | {{: | ||
| + | {{: | ||
| ---- | ---- | ||
| ==== FR-X ==== | ==== FR-X ==== | ||
| + | |||
| + | Switch ON the two fans of AFC-11. | ||
| + | |||
| + | {{: | ||
| + | |||
| + | Switch ON AFC-11 Goniometer controller. | ||
| + | |||
| + | {{: | ||
| + | |||
| + | Switch ON the vacuum pump below the Cobra Controller. | ||
| + | |||
| + | {{: | ||
| + | {{: | ||
| + | |||
| + | Switch ON the vacuum pump below the BioSAXS. | ||
| + | |||
| + | {{: | ||
| + | {{: | ||
| Switch ON the transformer. | Switch ON the transformer. | ||
| + | |||
| + | {{: | ||
| Reset and then switch ON the generator at the back. | Reset and then switch ON the generator at the back. | ||
| + | |||
| + | {{: | ||
| POWER ON on the front main panel. | POWER ON on the front main panel. | ||
| + | Press both DOOR LOCK buttons. | ||
| + | |||
| + | {{: | ||
| Start the Vacuum on the XG Control panel. | Start the Vacuum on the XG Control panel. | ||
| Wait until vacuum is below 1.10E-4 (about 1h30). | Wait until vacuum is below 1.10E-4 (about 1h30). | ||
| + | |||
| + | {{: | ||
| Turn X-Rays ON on the XG Control panel. | Turn X-Rays ON on the XG Control panel. | ||
| Line 39: | Line 74: | ||
| ==== Cryostream ==== | ==== Cryostream ==== | ||
| + | |||
| + | Switch ON the helium compressor. | ||
| + | |||
| + | {{: | ||
| + | {{: | ||
| Switch ON on the back of the device. | Switch ON on the back of the device. | ||
| + | |||
| + | {{: | ||
| Cycle Start/Start | Cycle Start/Start | ||
| For the first hour the temperature won't really decrease much. | For the first hour the temperature won't really decrease much. | ||
| Final temperature should be around 100 K. | Final temperature should be around 100 K. | ||
| + | |||
| + | {{: | ||
| ---- | ---- | ||
| Line 50: | Line 94: | ||
| ==== Detector ==== | ==== Detector ==== | ||
| + | Switch ON the EIGER water chiller. | ||
| + | |||
| + | {{: | ||
| + | |||
| + | Switch ON the power supply for the EIGER. | ||
| + | |||
| + | {{: | ||
| + | |||
| + | Switch ON the framegrabber. | ||
| + | |||
| + | {{: | ||
crystallography/collection/frx/on.1515054333.txt.gz · Last modified: (external edit)
