-==== Rigaku FR-X Usage ====+====== Rigaku FR-X Usage ======
  
 Training on 2015.03.30 Training on 2015.03.30
  
-FR-X generator: 45 kV, 66 mA, 70 μ cross-section, 180 μ beam, 1.7 miliradians divergence+**FR-X generator** 
 +  * 45 kV,  
 +  * 66 mA,  
 +  * 70 μ cross-section,  
 +  * 180 μ beam,  
 +  * 1.7 miliradians divergence
  
-Pilatus 300K detector: 3 chips, shorter exposure time, no readout (no shutter closing), almost never overloaded (>1.5 millions), water circulation at room temperature, dry air+**XG Control** tells you the settings values (45 kV and 66 mA vs 20 kV and 10 mA when not used) .  
 +You can screen as soon as the parameters have reached the target values, but wait one more hour to collect, so that the beam is stable. 
 + 
 +**Pilatus 300K detector** 
 +  * 3 chips,  
 +  * shorter exposure time,  
 +  * no readout (no shutter closing),  
 +  * almost never overloaded (>1.5 millions),  
 +  * water circulation at room temperature,  
 +  * dry air
  
 <code>>thread</code> command on camserver gives the temperature and humidity level <code>>thread</code> command on camserver gives the temperature and humidity level
Line 12: Line 26:
 Cryostream should be at 100 K Cryostream should be at 100 K
  
-Close 3 glass doors, push the green light on the left+Close 3 glass doors, push the green light on the left ((HKL does get info on the shutter status))
  
-Run HKL_3000_P300KChi (the other HKL_3000 version is for data integration only)+Run HKL_3000_P300KChi ((the other HKL_3000 version is for data integration only))
  
 <q>Collect/Connect to initialize</q> (also open an instrument server window in the background) <q>Collect/Connect to initialize</q> (also open an instrument server window in the background)
Line 27: Line 41:
 Set the directory /data/user/project1_crystal1 Set the directory /data/user/project1_crystal1
  
-**dtdisplay**+==== dtdisplay ==== 
   * middle click > zoom area   * middle click > zoom area
   * middle click > reset zoom   * middle click > reset zoom
Line 37: Line 52:
 ---- ----
  
-**Index**+==== Index ==== 
   * Refine in triclinic (P1)   * Refine in triclinic (P1)
   * Fit Basic   * Fit Basic
   * Fit All   * Fit All
   * Mosaicity (by default it cannot estimate well the mosaicity((default value 0.3)), so you need to select)   * Mosaicity (by default it cannot estimate well the mosaicity((default value 0.3)), so you need to select)
 +  * With a 3D Window >1, the mosaicity option with already be checked
   * To process up to the corner, you need to index up to the corner!   * To process up to the corner, you need to index up to the corner!
  
 ---- ----
  
-**Strategy**+==== Strategy ==== 
   * HKL_3000 strategy only works for single axis   * HKL_3000 strategy only works for single axis
   * Rigaku_Strategy is specially made for FR-X on the 4 axis   * Rigaku_Strategy is specially made for FR-X on the 4 axis
Line 61: Line 79:
 For some pins, Chi 60° may be too high and should be restricted to 50° For some pins, Chi 60° may be too high and should be restricted to 50°
  
-Close the window and choose <q>Setup DC</q> to transfer the informations to HKL data collection window+Close the window and choose <q>Setup DC</q> to transfer the information to HKL data collection window
 <wrap em>Exposure time is not imported from the strategy and should adjusted manually</wrap> <wrap em>Exposure time is not imported from the strategy and should adjusted manually</wrap>
  
 +multi mode is useful to find the best detector distance and exposure time
 +
 +Check diffraction images
 +go to the data directory
 +>dtdisplayhkl *.img &
 +
 +30 to 50 strong reflections are needed for indexing.
 +If it is not the case, increase the exposure time.
 +It's possible to index non consecutive images if they have the same scan id 
 +
 +At 2θ ω goes from -90° to +90°, so don't put 90° step with 0° as ω start
 +
 +Open display, do the peak search there, frame ↑ (middle click), the peak search has automatically changed, OK
 +
 +Chi geometry goniometer, omega angle 180° follows the 2θ angle
 +
 +If you press ABORT, the connexion will be lost and you will need to re-initialize
 +Check the log on the MSCservdect window.
 +To resume, check the parameters in the display.
 +The scan number will be implemented.
 +
 +With a 3D window >1, the mosaicity box will automatically be checked
 +
 +
 +----
 +
 +==== Zoom window ====
 +
 +
 +**Box size**: (16 by default) usually not necessary to change, even if there are neighbor spots in the box
 +Int_box: red ellipsoid to integrate spot. The reject area is between the red and the white. Is it necessary to increase the size for the red ellipsoid (0.35 by default)?
 +
 +Changing box size also changes the spot size. Use the box size that gives the highest resolution for the red area (it depends if the number is odd or even).
 +
 +**Profile fitting radius** is the yellow circle. When ON, it display only the strongest reflexions. Increase (from 8 to 10) when there are less spots to have enough strong reflexions within the circle, otherwise HKL might reject too many weak reflexions. Don't use a too big radius because spot profile is different at the center and in the edges.
 +
 +**3D window**: number of consecutive images put together to estimate the mosaicity. Usually it covers the spread of the mosaicity ONCE, but set it to cover it TWICE.
 +
 +**Set shadow region**: the Mask for the beamstop is not critical.
 +
 +
 +----
 +
 +==== Integrate ====
 + 
 +All scans are processed
 +χ² should be close to 1
 +
 +
 +----
 +
 +==== Scale ====
 +
 +**Check the boxes**:
 +  * Scale restrains
 +  * Absorption correction
 +  * Anomalous
 +  * B restrains (HKL recommended)
 +
 +Have low reflexions marked for rejection.
 +Scale several times until the number don't change anymore.
 +If χ² is too low, decrease the error scale factor.
 +Exclude frames if necessary.
 +
 +**For SAD phasing**:
 +if χ² is high: increase the scale factor (up to 1.6) and the error model (up to 0.6-0.7), but it might kill the anomalous signal. 
 +Check the orange(merged)/blue(unmerged) plot on χ² : higher signal means higher gap
 +
 +**Use auto correction**: will try to reach a χ² of 1, but we don't know how, so be careful.
 +
 +----
 +
 +==== Report ====
 +
 +The log file contains merged I+/I- an the end of the file and tables with unmerged I+/I- in the middle of the file.
 +
 +
 +----
 +
 +==== SAD-Phasing ====
 +
 +The sequence is necessary. Give the source or the signal or define if you have a heavy atom.
 +Software used: Arp/Warp, CCP4-6.3.0, CCP4-6.4.0(Coot), ShelX
 +
 +**Data analysis**: 
 +Grey area 1.3 for anomalous signal (can be used up to 1.0)
 +
 +**Find sites**:
 +Choose the resolution limit (>1.8 Å to distinguish disulfide bonds, test 0.1 Å increments)
 +Check the absolute value and the spread of solutions (a few in the up-right corner)
 +
 +**Site view**:
 +Sphere size correlates with the occupancy
 +
 +**Phase**:
 +High resolution limit; try different resolution cutoffs
 + 
 +**Check hand** (//ShelXE//):
 +You should see the difference, dark red/dark blue, difference of one graduation in scale.
 +If necessary, increase the cycle number so that a plateau is reached.
 +
 +**Phase**:
 +//Mlphase//, //DM-phase extension//
 +
 +**NCS**:
 +//DM// or //Parrot// (a bit better)
 +At least try it if you have more than 1 molecule in the Asymmetric Unit
 +
 +**Build**:
 +//ARP/WARP// for high resolution
 +//Buccaner// for lower resolution (∼ 3 Å)
 +Let ARP/WARP build/refine several times if has difficulties building. Start several times witha few cycles rather that once with many cycles.
 +
 +
 +----
 +
 +==== Molecular Replacement ====
 +
 +**Analyze data**
 +
 +**Prepare Model**: Edit the PDB file (delete ligands, waters, side-chains if you want)
 +
 +**Run MR** (//MolRep//): 3 Å by default; rigid body at 3.0 Å, then refine at max resolution
  
  