| Training on 2015.03.30 | Training on 2015.03.30 | ||
| - | FR-X generator: 45 kV, 66 mA, 70 μ cross-section, | + | **FR-X generator**: |
| + | * 45 kV, | ||
| + | * 66 mA, | ||
| + | * 70 μ cross-section, | ||
| + | * 180 μ beam, | ||
| + | * 1.7 miliradians divergence | ||
| **XG Control** tells you the settings values (45 kV and 66 mA vs 20 kV and 10 mA when not used) . | **XG Control** tells you the settings values (45 kV and 66 mA vs 20 kV and 10 mA when not used) . | ||
| You can screen as soon as the parameters have reached the target values, but wait one more hour to collect, so that the beam is stable. | You can screen as soon as the parameters have reached the target values, but wait one more hour to collect, so that the beam is stable. | ||
| - | Pilatus 300K detector: 3 chips, shorter exposure time, no readout (no shutter closing), almost never overloaded (>1.5 millions), water circulation at room temperature, | + | **Pilatus 300K detector**: |
| + | * 3 chips, | ||
| + | * shorter exposure time, | ||
| + | * no readout (no shutter closing), | ||
| + | * almost never overloaded (>1.5 millions), | ||
| + | * water circulation at room temperature, | ||
| + | * dry air | ||
| < | < | ||
| Line 168: | Line 177: | ||
| **Phase**: | **Phase**: | ||
| - | High resolution limit; try different resolution | + | High resolution limit; try different resolution |
| **Check hand** (// | **Check hand** (// | ||
| - | You should see the difference, | + | You should see the difference, |
| + | If necessary, increase the cycle number so that a plateau is reached. | ||
| **Phase**: | **Phase**: | ||
| Line 177: | Line 187: | ||
| **NCS**: | **NCS**: | ||
| + | //DM// or //Parrot// (a bit better) | ||
| At least try it if you have more than 1 molecule in the Asymmetric Unit | At least try it if you have more than 1 molecule in the Asymmetric Unit | ||
| Line 182: | Line 193: | ||
| // | // | ||
| // | // | ||
| + | Let ARP/WARP build/ | ||