 ====== Rigaku FR-X Usage ====== ====== Rigaku FR-X Usage ======
- 
- 
  
 Training on 2015.03.30 Training on 2015.03.30
  
-FR-X generator: 45 kV, 66 mA, 70 μ cross-section, 180 μ beam, 1.7 miliradians divergence+**FR-X generator** 
 +  * 45 kV,  
 +  * 66 mA,  
 +  * 70 μ cross-section,  
 +  * 180 μ beam,  
 +  * 1.7 miliradians divergence
  
 **XG Control** tells you the settings values (45 kV and 66 mA vs 20 kV and 10 mA when not used) .  **XG Control** tells you the settings values (45 kV and 66 mA vs 20 kV and 10 mA when not used) . 
 You can screen as soon as the parameters have reached the target values, but wait one more hour to collect, so that the beam is stable. You can screen as soon as the parameters have reached the target values, but wait one more hour to collect, so that the beam is stable.
  
-Pilatus 300K detector: 3 chips, shorter exposure time, no readout (no shutter closing), almost never overloaded (>1.5 millions), water circulation at room temperature, dry air+**Pilatus 300K detector** 
 +  * 3 chips,  
 +  * shorter exposure time,  
 +  * no readout (no shutter closing),  
 +  * almost never overloaded (>1.5 millions),  
 +  * water circulation at room temperature,  
 +  * dry air
  
 <code>>thread</code> command on camserver gives the temperature and humidity level <code>>thread</code> command on camserver gives the temperature and humidity level