User Tools

Site Tools


crystallography:collection:frx:usage

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
crystallography:collection:frx:usage [2016/07/12 09:35] bonnefoncrystallography:collection:frx:usage [2023/11/01 20:19] (current) – external edit 127.0.0.1
Line 3: Line 3:
 Training on 2015.03.30 Training on 2015.03.30
  
-FR-X generator: +**FR-X generator**
   * 45 kV,    * 45 kV, 
   * 66 mA,    * 66 mA, 
Line 13: Line 13:
 You can screen as soon as the parameters have reached the target values, but wait one more hour to collect, so that the beam is stable. You can screen as soon as the parameters have reached the target values, but wait one more hour to collect, so that the beam is stable.
  
-Pilatus 300K detector: 3 chips, shorter exposure time, no readout (no shutter closing), almost never overloaded (>1.5 millions), water circulation at room temperature, dry air+**Pilatus 300K detector** 
 +  * 3 chips,  
 +  * shorter exposure time,  
 +  * no readout (no shutter closing),  
 +  * almost never overloaded (>1.5 millions),  
 +  * water circulation at room temperature,  
 +  * dry air
  
 <code>>thread</code> command on camserver gives the temperature and humidity level <code>>thread</code> command on camserver gives the temperature and humidity level
crystallography/collection/frx/usage.1468316100.txt.gz · Last modified: (external edit)