 ===== Basic Microscope Alignment ===== ===== Basic Microscope Alignment =====
-THe microscope should be roughly aligned when you arrive.+The microscope should be roughly aligned when you arrive. 
 + 
 +__//**There is no need to do a complete alignment every session, if the microscope behaves strangely, ask for help!**//__ 
 + 
 +You should perform only two or three steps: 
 +  * Camera gain reference 
 +  * Direct alignment 
 +  * Objective Astigmatism / Coma-free alignment (OPTIONAL for screening!) 
 + 
 +==== Gain reference of the K2 camera ==== 
 +This is done in Digital Micrograph (DM) on the K2 server. 
 +  - Go to an empty region on the sample 
 +  - Go to high mag (e.g. 45000x) 
 +  - Insert 150µm C2 aperture 
 +  - Go to spot size 2 
 +  - Check on the FluScreen that beam is centered 
 +  - Digital Micrograph > Camera > Prepare Gain Reference 
 +  - Perform the gain in Linear mode first 
 +  - Adjust Beam Intensity to match the number of count on the camera 
 +  - Proceed to Counting mode 
 +  - Press 'Yes' to aquire the dark frame reference image (take twice 3 min 30) 
 +  - At the end of the dark ref update, you will be prompt to lower the dose: insert 30µm C2 aperture, go to spot size 6, adjust C2 lens to 39.6%) 
 +  - Lift the Fluscreen and aquire Counting Gain Reference Image (you DON'T need to match the number given by DM) 
 + 
 +==== Direct alignment ==== 
 +This is done in the microscope user interface. 
 + 
 +=== Prepare the microscope === 
 +  - Insert the calibration specimen (position 1 in the autoloader) 
 +  - Find an unbroken square 
 +  - Bring sample to eucentric high 
 +  - Press Eucentric Focus on the right control panel 
 +  - Set the microscope to the imaging conditions you want to use (e.g. mag 45000x, Spot size 6, C2 aperture 30µm) 
 +  - Perform direct alignment in microprobe and nanoprobe as follow: 
 + 
 +=== Microprobe === 
 +  - Switch the condenser system to microprobe 
 +  - Open Direct Alignment panel 
 +  - **DO NOT TOUCH GUN TILT/SHIFT** 
 +  - Perform Beam Tilt Pivot point X alignment 
 +  - Perform Beam Tilt Pivot point Y alignment 
 +  - Check C2 aperture centering 
 +  - Perform Rotation Center alignment 
 + 
 +=== Nanoprobe === 
 +  - Switch the condenser system to nanoprobe 
 +  - Open Direct Alignment panel 
 +  - Again, **DO NOT TOUCH GUN TILT/SHIFT** 
 +  - Redo Beam Tilt Pivot point X alignment 
 +  - Redo Beam Tilt Pivot point Y alignment 
 +  - Check C2 aperture centering 
 +  - Redo Rotation Center alignment 
 + 
 +Eventually, recenter the beam 
 + 
 +==== Objective Astigmatism / Coma-free alignment ==== 
 +This is done in SerialEM. You need to have a sample inserted to do it. Re-use the calibration specimen or a carbon Quantifoil (not a gold grid!). Use the Imaging state you will use for data collection (e.g. mag 45.000kx, spot 6, C2 aperture 30 µm, C2 lens 39.6%) 
 + 
 +  - Insert and center Objective Aperture 
 +  - In SerialEM, Focus/Tune > Correct Astigmatism by CTF 
 +  - In SerialEM, Focus/Tune > Coma-free by CTF
  