glacios_collect
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| glacios_collect [2022/02/11 15:18] – bonnefon | glacios_collect [2023/11/01 20:18] (current) – external edit 127.0.0.1 | ||
|---|---|---|---|
| Line 4: | Line 4: | ||
| ==== Beam microprobe and nanoprobe calibrations ==== | ==== Beam microprobe and nanoprobe calibrations ==== | ||
| + | ++++Prepare the Cross Grating grid for calibration| | ||
| * Introduce the Cross Grating Grid in the column. | * Introduce the Cross Grating Grid in the column. | ||
| * Go to x34 Imaging status. | * Go to x34 Imaging status. | ||
| Line 19: | Line 20: | ||
| * Push the **Wobbler** knob on the console. | * Push the **Wobbler** knob on the console. | ||
| * Adjust the Z height on the screen | * Adjust the Z height on the screen | ||
| + | ++++ | ||
| == Microprobe == | == Microprobe == | ||
| Line 92: | Line 94: | ||
| * In Record Tab, set 2 sec exposure, processing unnormalized, | * In Record Tab, set 2 sec exposure, processing unnormalized, | ||
| * If the signal is too weak, increase the time, concentrate the beam or change the carbon area | * If the signal is too weak, increase the time, concentrate the beam or change the carbon area | ||
| - | * ++++Correct astigmatism by CTF| | + | * ++Correct astigmatism by CTF|<div> |
| * Menu: Calibration > Focus & Tuning > Correct astigmatism by CTF | * Menu: Calibration > Focus & Tuning > Correct astigmatism by CTF | ||
| * Open FFT window with Shift+F | * Open FFT window with Shift+F | ||
| * Repeat several times until the corrected parameters have converged | * Repeat several times until the corrected parameters have converged | ||
| - | ++++ | + | </ |
| - | * ++++Coma free align by CTF| | + | * ++Coma free align by CTF|<div> |
| * Menu: Calibration > Focus & Tuning > Coma free align by CTF | * Menu: Calibration > Focus & Tuning > Coma free align by CTF | ||
| * Open FFT window with Shift+F | * Open FFT window with Shift+F | ||
| - | ++++ | + | </ |
| - | * ++++Coma vs image shift| | + | * ++Coma vs image shift|<div> |
| * Menu: Calibration > Focus & Tuning > Coma vs image shift | * Menu: Calibration > Focus & Tuning > Coma vs image shift | ||
| * Open FFT window with Shift+F | * Open FFT window with Shift+F | ||
| Line 107: | Line 109: | ||
| * parameters for R 1.2/1.3 grids: 5 μm | * parameters for R 1.2/1.3 grids: 5 μm | ||
| * parameters for R 2/2 grids: 6 μm | * parameters for R 2/2 grids: 6 μm | ||
| - | ++++ | + | </ |
| + | |||
| + | Introduce the Objective aperture and redo the Correct astigmatism by CTF step once. | ||
| + | |||
| + | ---- | ||
| + | |||
| + | ==== Configuration of the multihole / multishot acquisition ==== | ||
| + | |||
| + | Do this configuration just before starting the acquisition on one of the squares selected for the whole acquisition after correction of the image shift in x280 and LD mode imaging states. | ||
| + | |||
| + | * On the square map, mark the positions of the 4 holes at the corners of a 9 holes square. | ||
| + | * Go to the center hole with **Go to marker** and properly center on the hole with **Stage shift**. | ||
| + | * Press View in LD mode (ensure that defocus is below -20) | ||
| + | * Open Menu: // | ||
| + | * Change the parameters of the multishot for calibration | ||
| + | * Click on the central hole for positioning | ||
| + | * Adjust the number of shots per hole (in general 3) | ||
| + | * Adjust the distance (72 in general for R 2/2 grids) | ||
| + | * Menu: // | ||
| + | * Select regular pattern 3 x 3 | ||
| + | * ++Calibrate after ensuring proper centering on the center hole|< | ||
| + | * Move by Image shift to the first corner | ||
| + | * Center properly in the hole using zoom if necessary | ||
| + | * Save the position by clicking on **Save image shift** | ||
| + | * Proceed to the next corner and repeat | ||
| + | </ | ||
| + | * Save the settings in //Menu > Settings > Save// | ||
| + | |||
| + | == Beam tuning == | ||
| + | |||
| + | * Ensure that LD modes View and Record are perfectly aligned | ||
| + | * Calibrate the proper centering in Record/ | ||
| + | |||
| + | * Precisely center Record beam with Direct Alignments/ | ||
| + | * Check **Continuous update** | ||
| + | * Press View and recenter if necessary with set add beam shift | ||
| + | * ++Then cycle several times between the modes until the beam stop moving|< | ||
| + | * Press **Record** and correct Direct Alignments/ | ||
| + | * Press **Focus** and set add beam shift | ||
| + | * Press **Trial** and set add beam shift | ||
| + | * Repeat | ||
| + | </ | ||
| + | * Uncheck **Continuous update** | ||
glacios_collect.1644592720.txt.gz · Last modified: (external edit)