glacios_collect [2022/02/11 15:43] bonnefonglacios_collect [2023/11/01 20:18] (current) – external edit 127.0.0.1
Line 124: Line 124:
   * Open Menu: //Navigation > Montaging & grids > Set multi//   * Open Menu: //Navigation > Montaging & grids > Set multi//
   * Change the parameters of the multishot for calibration   * Change the parameters of the multishot for calibration
-  Click on the central hole for positioning+  Click on the central hole for positioning
   * Adjust the number of shots per hole (in general 3)   * Adjust the number of shots per hole (in general 3)
   * Adjust the distance (72 in general for R 2/2 grids)   * Adjust the distance (72 in general for R 2/2 grids)
   * Menu: //Navigation > Montaging & grids > Find hole in regul// to configure   * Menu: //Navigation > Montaging & grids > Find hole in regul// to configure
-  Select regular pattern 3 x 3+  Select regular pattern 3 x 3
   * ++Calibrate after ensuring proper centering on the center hole|<div>   * ++Calibrate after ensuring proper centering on the center hole|<div>
   * Move by Image shift to the first corner   * Move by Image shift to the first corner
Line 134: Line 134:
   * Save the position by clicking on **Save image shift**   * Save the position by clicking on **Save image shift**
   * Proceed to the next corner and repeat   * Proceed to the next corner and repeat
-</div>+</div>++ 
 +  * Save the settings in //Menu > Settings > Save// 
 + 
 +== Beam tuning == 
 + 
 +  * Ensure that LD modes View and Record are perfectly aligned 
 +  * Calibrate the proper centering in Record/View/Focus/Trial modes 
 +  
 +  * Precisely center Record beam with Direct Alignments/Beam shift 
 +  * Check **Continuous update** 
 +  * Press View and recenter if necessary with set add beam shift 
 +  * ++Then cycle several times between the modes until the beam stop moving|<div> 
 +  * Press **Record** and correct Direct Alignments/Beam shift 
 +  * Press **Focus** and set add beam shift 
 +  * Press **Trial** and set add beam shift 
 +  * Repeat 
 +</div>++ 
 +  * Uncheck **Continuous update**
  
glacios_collect.1644594180.txt.gz · Last modified: (external edit)