User Tools

Site Tools


glacios_interface

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
glacios_interface [2023/03/20 16:11] crucifixglacios_interface [2023/11/01 20:18] (current) – external edit 127.0.0.1
Line 1: Line 1:
-====== The TF20 microscope interfaces ======+====== The Glacios microscope interfaces ======
  
  
 ===== The microscope console ===== ===== The microscope console =====
  
-**Location **: right in front of you on the desk (screen on the left)+**Location**: right in front of you on the desk
  
-**Role**: TEM User Interface provides direct control of some of the most important microscope parts:+**Role**: it provides direct control of some of the most important microscope parts:
   * Magnification   * Magnification
   * Electron dose: spot size (C1 lens) and beam spread (C2 lens)   * Electron dose: spot size (C1 lens) and beam spread (C2 lens)
Line 12: Line 12:
   * Stage position (with the joystick and Z buttons)   * Stage position (with the joystick and Z buttons)
   * Beam position (with the trackball)   * Beam position (with the trackball)
-  * and many other actions through the programmable switches (L/R1-2-3) and Multifunction X/Y knobs. +  * and many other actions through the programmable switches (L/R1-2-3) and Multifunction X/Y knobs. 
- +
-* ++For more details see:|<div> +
- +
-  * {{ :capture_tui_description.jpg?nolink&600 |}}+
  
 ---- ----
Line 24: Line 20:
 **Location**: On the left screen **Location**: On the left screen
  
-**Role**: It controls the US10001 Camera.+**Role**: It controls the K2 direct electron detector.
  
-  * During screening, DM will allow acquisition in continuous mode+  * The beginner user will not use it
- +  * ++The expert user will use it to:|<div> 
 +  * prepare Dark, linear and counting / Super Resolution gain references. 
 +  * measure the dose. 
 +  * monitor the K2 health status. 
 +  * troubleshoot the system. 
 +  * perform cryo-cycle of the detector 
 +</div>++
  
 ---- ----
Line 33: Line 35:
 ===== SerialEM User Interface (SEM) ===== ===== SerialEM User Interface (SEM) =====
  
-**Location**: On the right screen screen+**Location**: On the left screen
  
 **Role**: It provides camera/microscope commands to allow the user to screen its grids and setup an automated data collection. **Role**: It provides camera/microscope commands to allow the user to screen its grids and setup an automated data collection.
  
-  * ++The user will use it to:|<div>+  * ++The beginner user will use it to:|<div>
   * record the Atlases.   * record the Atlases.
   * select squares and prepare the square maps at eucentric height.   * select squares and prepare the square maps at eucentric height.
   * select targets (holes).   * select targets (holes).
   * screen targets (holes).   * screen targets (holes).
-  * center the different beam modes all together on the camera axis. 
 </div>++ </div>++
   * ++The expert user will use it to:|<div>   * ++The expert user will use it to:|<div>
Line 48: Line 49:
   * perform beam corrections (astigmatism, coma-free).   * perform beam corrections (astigmatism, coma-free).
   * perform calibrations.   * perform calibrations.
 +  * setup the dose rate, total dose and dose fractionning.
 +  * setup the time and imaging modes for the different beam modes.
 +  * center the different beam modes all together on the camera axis.
 </div>++ </div>++
  
glacios_interface.1679328708.txt.gz · Last modified: (external edit)