glacios_screening [2023/01/17 13:54] – [6. Prepare Serial-EM for screening] marechanglacios_screening [2023/11/01 20:18] (current) – external edit 127.0.0.1
Line 4: Line 4:
  
  
-==== 1. Load the Grids ====+==== 1. Load  Grid ====
  
   - ++Introduce your grids in the cassette|<div>   - ++Introduce your grids in the cassette|<div>
Line 13: Line 13:
   - ++Check Autoloder temperatures, then load the NanoCab in the microscope|<div>{{ :autoloader_tempertures_all.png?nolink |}}   - ++Check Autoloder temperatures, then load the NanoCab in the microscope|<div>{{ :autoloader_tempertures_all.png?nolink |}}
 </div>++ </div>++
-  - Remove the NanoCab+ 
 + 
 + 
  
  
Line 33: Line 36:
 </div>++ </div>++
   - ++Load platform or user settings|<div>   - ++Load platform or user settings|<div>
-  * Menu: //Settings > Open > SerialEM_Settings_SPA//+  * Menu: //Settings > Open > Platform_Settings_SPA//
 </div>++  </div>++ 
  
Line 239: Line 242:
   - With **multifunction X & Y knobs**, center the beam on the fluorescent screen, then press //Done//.   - With **multifunction X & Y knobs**, center the beam on the fluorescent screen, then press //Done//.
   - Switch to **View beam** with the //Vie.// button. Tick //Set// in //Additional beam shift//. **View beam only need to be centered once**.   - Switch to **View beam** with the //Vie.// button. Tick //Set// in //Additional beam shift//. **View beam only need to be centered once**.
-  - With the **trackball**, center the view beam on the fluorescent screen.+  - With the **trackball**, center the view beam on the fluorescent screen. **Untick** the //Set// button.
   - Center the Record beam again as described above.   - Center the Record beam again as described above.
   - Switch to **Focus beam** (//Foc.// button) and change its diameter with the intensity knob according with the desired focus surface (depends on your hole spacing).   - Switch to **Focus beam** (//Foc.// button) and change its diameter with the intensity knob according with the desired focus surface (depends on your hole spacing).