===== Turn in-house FR-X OFF ===== ---- ==== Detector ==== Open Firefox and connect to 192.168.127.50 {{:crystallography:collection:frx:framegrabber_1.png?600|}} Select 'system settings'. Username = eiger, PW = #EIGER_Detector# In the lefthand column select 'EIGER' and then 'DCU-Control' {{:crystallography:collection:frx:framegrabber_2.png?600|}} select "Power OFF" to shut down the framegrabber. {{:crystallography:collection:frx:framegrabber_3.png?600|}} Switch OFF the power supply for the EIGER. {{:crystallography:collection:frx:eiger_power_supply.jpg?600|}} Switch OFF the EIGER water chiller. {{:crystallography:collection:frx:eiger_chiller.jpg?400|}} ---- ==== FR-X ==== Decrease gradually the the current 66 to 10 mA and then voltage from 45 kV to 20 kV. First step 66->60 mA, (45->40 kV). Wait 10 s. Then decrease by steps of 10. (Lower the current first.) {{:crystallography:collection:frx:xg.png?600|}} Turn X-RAY OFF on the XG Control panel. Wait 1h. Stop the Vacuum on the XG Control panel. Wait 1h. POWER OFF on the FR-X main panel. {{:crystallography:collection:frx:frx_front.jpg?600|}} Switch OFF at the back of the FR-X. {{:crystallography:collection:frx:frx_back.jpg?400|}} Switch the main power source OFF. {{:crystallography:collection:frx:main_power_supply.jpg?600|}} Switch OFF AFC-11 Goniometer controller. {{:crystallography:collection:frx:afc11_closeup.jpg?600|}} Switch OFF the two fans of AFC-11. {{:crystallography:collection:frx:acf11_back.jpg?400|}} Switch OFF the vacuum pump below the Cobra Controller. {{:crystallography:collection:frx:pump1.jpg?250|}} {{:crystallography:collection:frx:pump1_closeup.jpg?250|}} Switch OFF the vacuum pump below the BioSAXS. {{:crystallography:collection:frx:pump2.jpg?250|}} {{:crystallography:collection:frx:pump2_closeup.jpg?350|}} ---- ==== Cryostream ==== On the Cobra Controller : {{:crystallography:collection:frx:cobra_front.jpg?600|}} * Program / Ramp (change with dial) / Enter * Choose the speed (up to 360K/h) / Enter * Target / 293 K (Ambient) / Enter * Start * When ambient temperature is reached, press Stop Switch OFF at the back. {{:crystallography:collection:frx:cobra_back.jpg?600|}} Switch OFF the helium compressor. {{:crystallography:collection:frx:helium.jpg?300|}} {{:crystallography:collection:frx:helium_back.jpg?250|}} ---- ==== Technical Room E0044 ==== {{:crystallography:collection:frx:technical_room.jpg?400|}} Switch OFF the nitrogen compressor using the touch panel. Switch OFF the compressor on the back of the device. {{:crystallography:collection:frx:n2.jpg?150|}} {{:crystallography:collection:frx:n2_closeup.jpg?400|}} Switch OFF the main switch of the water cooling device (Haskins). {{:crystallography:collection:frx:haskins.jpg?200|}} {{:crystallography:collection:frx:haskins_closeup.jpg?400|}} Close the water supply. (red valve) {{:crystallography:collection:frx:red_valve.jpg?400|}} ---- ==== EIGER Computer ==== Shut down the computer. {{:crystallography:collection:frx:pc.jpg?600|}} ---- ==== BioSAXS ==== Switch OFF {{:crystallography:collection:frx:saxs_control_1.jpg?600|}} Switch OFF at the back. {{:crystallography:collection:frx:saxs_control_2.jpg?600|}} Remove the small middle screw to access the Pilatus 100K detector and switch OFF. {{:crystallography:collection:frx:pilatus100k_cover.jpg?200|}} {{:crystallography:collection:frx:pilatus100k.jpg?400|}} ---- ==== BioSAXS Computers ==== Shut down the SAXS PC. {{:crystallography:collection:frx:saxs_pc.jpg?600|}} Switch the display and control to the framegrabber with the Switch. {{:crystallography:collection:frx:saxs_pc_switch.jpg?400|}} Shut down the SAXS framegrabber.