====== Setup an automated data collection at Glacios ====== ==== Beam microprobe and nanoprobe calibrations ==== ++++Prepare the Cross Grating grid for calibration| * Introduce the Cross Grating Grid in the column. * Go to x34 Imaging status. * Select 2 positions: a carbon square and an empty square. * Introduce the Condenser 2 of 30 um * Go to x280 Imaging status. * In DigitalMicrograph click View and place the mouse arrow on a feature. * Use Alpha Toogle 20° in Stage² window (TEM) to rock the grid. * Correct with **Z axis knobs** on the console. * Repeat a second time. * Go to x45k Imaging status * In Low Dose Control window (SEM) click on Go to: Record. * Push the **Eucentric Focus** knob on the console. * Inset the screen (FC). * Push the **Wobbler** knob on the console. * Adjust the Z height on the screen with the **Z axis knobs** on the console. ++++ == Microprobe == Focus the beam and center it (with the trackball). * Go to microprobe (TEM) * Push the **Eucentric Focus** knob on the console. * In the **Direct Alignments** window: - Adjust the ++Beam tilt pp X|
* Click on **Beam tilt pp X** * Adjust with **Multifunction X/Y** knobs * Click **Done** at the bottom of the window.
++ - Adjust the ++Beam tilt pp Y|
* Click on **Beam tilt pp Y**. * Adjust with **Multifunction X/Y** knobs. * Click **Done** at the bottom of the window.
++ - Adjust the ++Beam shift|
* Click on **Beam shift**. * Adjust with **Multifunction X/Y** knobs. * Click **Done** at the bottom of the window.
++ - Adjust the ++Rotation center|
* Click on **Rotation center**. * Adjust with **Multifunction X/Y** knobs. * Click **Done** at the bottom of the window.
++ == Nanoprobe == Focus the beam and center it (with the trackball). * Go to nanoprobe (TEM) * Push the **Eucentric Focus** knob on the console. * Check the Condenser 2 centering and adjust if necessary. * In the **Direct Alignments** window: - Adjust the ++Beam tilt pp X|
* Click on **Beam tilt pp X** * Adjust with **Multifunction X/Y** knobs * Click **Done** at the bottom of the window
++ - Adjust the ++Beam tilt pp Y|
* Click on **Beam tilt pp Y** * Adjust with **Multifunction X/Y** knobs. * Click **Done** at the bottom of the window
++ - Adjust the ++Beam shift|
* Click on **Beam shift** * Adjust with **Multifunction X/Y** knobs * Click **Done** at the bottom of the window
++ - Adjust the ++Rotation center|
* Click on **Rotation center** * Adjust with **Multifunction X/Y** knobs * Click **Done** at the bottom of the window
++ ++++After all calibrations save the settings| * Menu: //Settings > Save// * In **Low Dose Control** window check **Continuous update** * Click on Update it * Uncheck **Continuous update** * Menu: //Settings > Save// ++++ ==== Correct astigmatism ==== * Verify that the **Turbo Auto Off** Vacuum option is checked (TEM) * Go to x45k Imaging status * In the Camera and script window (SEM) open the **Setup** * In Record Tab, set 2 sec exposure, processing unnormalized, align frames, save frames and the objective diaphragm out * If the signal is too weak, increase the time, concentrate the beam or change the carbon area * ++Correct astigmatism by CTF|
* Menu: Calibration > Focus & Tuning > Correct astigmatism by CTF * Open FFT window with Shift+F * Repeat several times until the corrected parameters have converged
++ * ++Coma free align by CTF|
* Menu: Calibration > Focus & Tuning > Coma free align by CTF * Open FFT window with Shift+F
++ * ++Coma vs image shift|
* Menu: Calibration > Focus & Tuning > Coma vs image shift * Open FFT window with Shift+F * If the signal is too weak, increase the time, set binning, concentrate the beam or change the carbon area * parameters for R 1.2/1.3 grids: 5 μm * parameters for R 2/2 grids: 6 μm
++ Introduce the Objective aperture and redo the Correct astigmatism by CTF step once. ---- ==== Configuration of the multihole / multishot acquisition ==== Do this configuration just before starting the acquisition on one of the squares selected for the whole acquisition after correction of the image shift in x280 and LD mode imaging states. * On the square map, mark the positions of the 4 holes at the corners of a 9 holes square. * Go to the center hole with **Go to marker** and properly center on the hole with **Stage shift**. * Press View in LD mode (ensure that defocus is below -20) * Open Menu: //Navigation > Montaging & grids > Set multi// * Change the parameters of the multishot for calibration * Click on the central hole for positioning * Adjust the number of shots per hole (in general 3) * Adjust the distance (72 in general for R 2/2 grids) * Menu: //Navigation > Montaging & grids > Find hole in regul// to configure * Select regular pattern 3 x 3 * ++Calibrate after ensuring proper centering on the center hole|
* Move by Image shift to the first corner * Center properly in the hole using zoom if necessary * Save the position by clicking on **Save image shift** * Proceed to the next corner and repeat
++ * Save the settings in //Menu > Settings > Save// == Beam tuning == * Ensure that LD modes View and Record are perfectly aligned * Calibrate the proper centering in Record/View/Focus/Trial modes * Precisely center Record beam with Direct Alignments/Beam shift * Check **Continuous update** * Press View and recenter if necessary with set add beam shift * ++Then cycle several times between the modes until the beam stop moving|
* Press **Record** and correct Direct Alignments/Beam shift * Press **Focus** and set add beam shift * Press **Trial** and set add beam shift * Repeat
++ * Uncheck **Continuous update**