==== Make sure Microscope is in good shape (ppt associated) ====
*++ a.Make sure HT and FEG are on|
* On TEM User Interface (TUI)
* setup tab > High Tension panel High tension button is highlighted yellow
* setup tab > FEG Control panel Operate button is highlighted yellow
{{ :capture_ht-operate-on.jpg?nolink |}}
++
*++ b.Make sure column valves are closed |
* On TEM User Interface (TUI)
* setup tab > Vaccum panel Col. Valves closed button is highlighted yellow
IMAGE
++
*++ c.Make sure that EM column vaccum is stable (IGP1 = 6)|
* On TEM User Interface (TUI)
* setup tab > Vaccum panel Column value indicates 6
IMAGE
++
*++ d.Make sure stage is at neutral position |
* On TEM User Interface (TUI)
* Stage2 tab > control panel press X,Y button to zero out the X, Y to center the stage
IMAGE
++
- ++Setup the microscope for recording the Atlases|
* On TEM user interface (**TUI**).
++