Clipping Procedure

source : https://www.jonipsaro.com/post/krios/

Before starting

Load C-clips into the holder wands.

Prepare the transfer station setup

  Arrange the transfer setup as pictured. It should include the specimen transfer pedestal, loaded C-clip wands, and autoloader forceps.
  Additionally, place clip rings (flat-side down) in each of the four mounting positions in the brass portion of the transfer pedestal.
  Inspect each clip ring to make sure that the clip side is face-up.
  Once clip rings are in place, rotate the brass disc such that all clips are covered. This will ensure they do not move while cooling the station.
  Place cryo-EM grid box(es) on the sample pedestal. Open them using the grid box fastener wand. Keep the grid box top attached to the fastner wand until needed.