Basic Microscope Alignment
The microscope should be roughly aligned when you arrive.
There is no need to do a complete alignment every session, if the microscope behaves strangely, ask for help!
You should perform only two or three steps:
Gain reference of the K2 camera
This is done in Digital Micrograph (DM) on the K2 server.
Go to an empty region on the sample
Go to high mag (e.g. 45000x)
Insert 150µm C2 aperture
Go to spot size 2
Check on the FluScreen that beam is centered
Digital Micrograph > Camera > Prepare Gain Reference
Perform the gain in Linear mode first
Adjust Beam Intensity to match the number of count on the camera
Proceed to Counting mode
Press 'Yes' to aquire the dark frame reference image (take twice 3 min 30)
At the end of the dark ref update, you will be prompt to lower the dose: insert 30µm C2 aperture, go to spot size 6, adjust C2 lens to 39.6%)
Lift the Fluscreen and aquire Counting Gain Reference Image (you DON'T need to match the number given by DM)
Direct alignment
This is done in the microscope user interface.
Prepare the microscope
Insert the calibration specimen (position 1 in the autoloader)
Find an unbroken square
Bring sample to eucentric high
Press Eucentric Focus on the right control panel
Set the microscope to the imaging conditions you want to use (e.g. mag 45000x, Spot size 6, C2 aperture 30µm)
Perform direct alignment in microprobe and nanoprobe as follow:
Microprobe
Switch the condenser system to microprobe
Open Direct Alignment panel
DO NOT TOUCH GUN TILT/SHIFT
Perform Beam Tilt Pivot point X alignment
Perform Beam Tilt Pivot point Y alignment
Check C2 aperture centering
Perform Rotation Center alignment
Nanoprobe
Switch the condenser system to nanoprobe
Open Direct Alignment panel
Again, DO NOT TOUCH GUN TILT/SHIFT
Redo Beam Tilt Pivot point X alignment
Redo Beam Tilt Pivot point Y alignment
Check C2 aperture centering
Redo Rotation Center alignment
Eventually, recenter the beam
Objective Astigmatism / Coma-free alignment
This is done in SerialEM. You need to have a sample inserted to do it. Re-use the calibration specimen or a carbon Quantifoil (not a gold grid!). Use the Imaging state you will use for data collection (e.g. mag 45.000kx, spot 6, C2 aperture 30 µm, C2 lens 39.6%)
Insert and center Objective Aperture
In SerialEM, Focus/Tune > Correct Astigmatism by CTF
In SerialEM, Focus/Tune > Coma-free by CTF