Beam microprobe and nanoprobe calibrations

Prepare the Cross Grating grid for calibration

Microprobe

Focus the beam and center it (with the trackball).

  1. Adjust the Beam tilt pp X
  2. Adjust the Beam tilt pp Y
  3. Adjust the Beam shift
  4. Adjust the Rotation center
Nanoprobe

Focus the beam and center it (with the trackball).

  1. Adjust the Beam tilt pp X
  2. Adjust the Beam tilt pp Y
  3. Adjust the Beam shift
  4. Adjust the Rotation center

After all calibrations save the settings

Correct astigmatism

Introduce the Objective aperture and redo the Correct astigmatism by CTF step once.


Configuration of the multihole / multishot acquisition

Do this configuration just before starting the acquisition on one of the squares selected for the whole acquisition after correction of the image shift in x280 and LD mode imaging states.

Beam tuning