Turn in-house FR-X OFF
Detector
Open firefox and connect to 192.168.127.50
Select system settings. Username = eiger, PW = #EIGER_Detector#
In the lefthand column select 'EIGER' and then 'DCU-Control'
select “Power OFF” to shut down the framegrabber.
Switch OFF the power supply for the EIGER.
Switch OFF the EIGER water chiller.
FR-X
Decrease gradually the the current 66 to 10 mA and then voltage from 45 kV to 20 kV. First step 66→60 mA, (45→40 kV). Wait 10 s. Then decrease by steps of 10. (Lower the current first)
Turn X-RAY OFF on the XG Control panel. Wait 1h.
Stop the Vacuum on the XG Control panel. Wait 1h.
POWER OFF on the FR-X main panel.
Switch OFF at the back of the FR-X.
Switch the main power source OFF.
Switch OFF AFC-11 Goniometer controller.
Switch OFF the two fans of AFC-11.
Switch OFF vacuum pump below Cobra
Switch OFF vacuum pump below BioSAXS
Cryostream
Program / Ramp (change with dial) / Enter
Choose the speed (up to 360K/h) / Enter
Target / Ambient / Enter
Enter / Enter
Start
When ambient temperature is reached, press Stop
Switch OFF at the back
Switch off the helium compressor