Rigaku FR-X Usage
Training on 2015.03.30
FR-X generator: 45 kV, 66 mA, 70 μ cross-section, 180 μ beam, 1.7 miliradians divergence
Pilatus 300K detector: 3 chips, shorter exposure time, no readout (no shutter closing), almost never overloaded (>1.5 millions), water circulation at room temperature, dry air
>thread
command on camserver gives the temperature and humidity level Green light at the back of the detector to check before use.
Cryostream should be at 100 K
Close 3 glass doors, push the green light on the left
Run HKL_3000_P300KChi (the other HKL_3000 version is for data integration only)
Collect/Connect to initialize
(also open an instrument server window in the background)
0.25° oscillation by default As a “rule of thumb” set the detector distance (in mm) to half the longest unit cell parameter (in Å) 1)
Align/Zero goniostat
resets all positions (parameters in /hkl_dc.ini)
2θ always swings away (clockwise)
Set the directory /data/user/project1_crystal1
dtdisplay
- middle click > zoom area
- middle click > reset zoom
- control + left click > move around
- measure /Measure Left 1x, 2x, 3x or 4x (to average)
Never collect at 2θ = 0, use at least 2θ = 2 to offset the center horizontally and avoid thus avoid missing data due to the gaps between detectors
Index
- Refine in triclinic (P1)
- Fit Basic
- Fit All
- Mosaicity (by default it cannot estimate well the mosaicity2), so you need to select)
- To process up to the corner, you need to index up to the corner!
Strategy
- HKL_3000 strategy only works for single axis
- Rigaku_Strategy is specially made for FR-X on the 4 axis
- Choose I want to make changes
- Completeness 99% (to reduce collect time)
- Redundancy 3.0 (4.0 by default)
- Distance (the distance from the first image is set)
- Suggested distance is not useful
- Edit preferences is possible
You can swing the slider on Time window to adjust/reduce the number of collected frames Don't forget to adjust the Mosaicity value! If anomalous data is required, adjust the redundancy accordingly
For some pins, Chi 60° may be too high and should be restricted to 50°
Close the window and choose Setup DC
to transfer the informations to HKL data collection window
Exposure time is not imported from the strategy and should adjusted manually