glacios_alignment
This is an old revision of the document!
Table of Contents
Basic Microscope Alignment
The microscope should be roughly aligned when you arrive.
There is no need to do a complete alignment every session, if the microscope behaves strangely, ask for help!
You should perform only two or three steps:
- Camera gain reference
- Direct alignment
- Objective Astigmatism / Coma-free alignment (OPTIONAL for screening!)
Gain reference of the K2 camera
This is done in Digital Micrograph (DM) on the K2 server.
- Go to an empty region on the sample
- Go to high mag (e.g. 45000x)
- Insert 150µm C2 aperture
- Go to spot size 2
- Check on the FluScreen that beam is centered
- Digital Micrograph > Camera > Prepare Gain Reference
- Perform the gain in Linear mode first
- Adjust Beam Intensity to match the number of count on the camera
- Proceed to Counting mode
- Press 'Yes' to aquire the dark frame reference image (take twice 3 min 30)
- At the end of the dark ref update, you will be prompt to lower the dose: insert 30µm C2 aperture, go to spot size 6, adjust C2 lens to 39.6%)
- Lift the Fluscreen and aquire Counting Gain Reference Image (you DON'T need to match the number given by DM)
Direct alignment
This is done in microscope interface.
- Insert calibration specimen (position 1 in the autoloader)
- Find an unbroken square
- Bring sample to eucentric high
- Press Eucentric Focus on the right control panel
Open the Direct alignment tab
DO NOT TOUCH GUN TILT/SHIFT
Objective Astigmatism / Coma-free alignment
This is done in SerialEM.
glacios_alignment.1668776957.txt.gz · Last modified: (external edit)
