Setup an automated data collection at Glacios
Beam microprobe and nanoprobe calibrations
Introduce the Cross Grating Grid in the column. Go to x34 Imaging status. Select 2 positions: a carbon square and an empty square. Introduce the Condenser 2 of 30 um Go to x280 Imaging status. In DigitalMicrograph click View and place the mouse arrow on a feature. Use Alpha Toogle 20° in Stage² window (TEM) to rock the grid. Correct with Z axis knobs on the console. Repeat a second time. Go to x45k Imaging status In Low Dose Control window (SEM) click on Go to: Record. Push the Eucentric Focus knob on the console.