Setup an automated data collection at Glacios

Beam microprobe and nanoprobe calibrations

  • Introduce the Cross Grating Grid in the column.
  • Go to x34 Imaging status.
  • Select 2 positions: a carbon square and an empty square.
  • Introduce the Condenser 2 of 30 um
  • Go to x280 Imaging status.
  • In DigitalMicrograph click View and place the mouse arrow on a feature.
  • Use Alpha Toogle 20° in Stage² window (TEM) to rock the grid.
  • Correct with Z axis knobs on the console.
  • Repeat a second time.
  • Go to x45k Imaging status
  • In Low Dose Control window (SEM) click on Go to: Record.
  • Push the Eucentric Focus knob on the console.
glacios_collect.1644590172.txt.gz · Last modified: (external edit)