Setup an automated data collection at Glacios

Beam microprobe and nanoprobe calibrations

  • Introduce the Cross Grating Grid in the column.
  • Go to x34 Imaging status.
  • Select 2 positions: a carbon square and an empty square.
  • Introduce the Condenser 2 of 30 um
  • Go to x280 Imaging status.
  • In DigitalMicrograph click View and place the mouse arrow on a feature.
  • Use Alpha Toogle 20° in Stage² window (TEM) to rock the grid.
  • Correct with Z axis knobs on the console.
  • Repeat a second time.
  • Go to x45k Imaging status
  • In Low Dose Control window (SEM) click on Go to: Record.
  • Push the Eucentric Focus knob on the console.
  • Inset the screen (FC).
  • Push the Wobbler knob on the console.
  • Adjust the Z height on the screen with the Z axis knobs on the console.

Microprobe

Focus the beam and center it (with the trackball).

  • Go to microprobe (TEM)
  • Push the Eucentric Focus knob on the console.
  • In the Direct Alignments window:
  1. Adjust the Beam tilt pp X
  2. Adjust the Beam tilt pp Y
  3. Adjust the Beam shift
  4. Adjust the Rotation center

Nanoprobe

Focus the beam and center it (with the trackball).

  • Go to nanoprobe (TEM)
  • Push the Eucentric Focus knob on the console.
  • Check the Condenser 2 centering and adjust if necessary.
  • In the Direct Alignments window:
  1. Adjust the Beam tilt pp X
  2. Adjust the Beam tilt pp Y
  3. Adjust the Beam shift
  4. Adjust the Rotation center

++++After all calibrations save the settings|

  • Menu: Settings > Save
  • In Low Dose Control window check Continuous update
  • Click on Update it
  • Unchek Continuous update