Setup an automated data collection at Glacios
Beam microprobe and nanoprobe calibrations
- Introduce the Cross Grating Grid in the column.
- Go to x34 Imaging status.
- Select 2 positions: a carbon square and an empty square.
- Introduce the Condenser 2 of 30 um
- Go to x280 Imaging status.
- In DigitalMicrograph click View and place the mouse arrow on a feature.
- Use Alpha Toogle 20° in Stage² window (TEM) to rock the grid.
- Correct with Z axis knobs on the console.
- Repeat a second time.
- Go to x45k Imaging status
- In Low Dose Control window (SEM) click on Go to: Record.
- Push the Eucentric Focus knob on the console.
- Inset the screen (FC).
- Push the Wobbler knob on the console.
- Adjust the Z height on the screen with the Z axis knobs on the console.
Microprobe
Focus the beam and center it (with the trackball).
- Go to microprobe (TEM)
- Push the Eucentric Focus knob on the console.
- In the Direct Alignments window:
- Adjust the Beam tilt pp X
- Adjust the Beam tilt pp Y
- Adjust the Beam shift
- Adjust the Rotation center
Nanoprobe
Focus the beam and center it (with the trackball).
- Go to nanoprobe (TEM)
- Push the Eucentric Focus knob on the console.
- Check the Condenser 2 centering and adjust if necessary.
- In the Direct Alignments window:
- Adjust the Beam tilt pp X
- Adjust the Beam tilt pp Y
- Adjust the Beam shift
- Adjust the Rotation center