Setup an automated data collection at Glacios
Beam microprobe and nanoprobe calibrations
Microprobe
Focus the beam and center it (with the trackball).
- Go to microprobe (TEM)
- Push the Eucentric Focus knob on the console.
- In the Direct Alignments window:
- Adjust the Beam tilt pp X
- Adjust the Beam tilt pp Y
- Adjust the Beam shift
- Adjust the Rotation center
Nanoprobe
Focus the beam and center it (with the trackball).
- Go to nanoprobe (TEM)
- Push the Eucentric Focus knob on the console.
- Check the Condenser 2 centering and adjust if necessary.
- In the Direct Alignments window:
- Adjust the Beam tilt pp X
- Adjust the Beam tilt pp Y
- Adjust the Beam shift
- Adjust the Rotation center
Correct astigmatism
- Verify that the Turbo Auto Off Vacuum option is checked (TEM)
- Go to x45k Imaging status
- In the Camera and script window (SEM) open the Setup
- In Record Tab, set 2 sec exposure, processing unnormalized, align frames, save frames and the objective diaphragm out
- If the signal is too weak, increase the time, concentrate the beam or change the carbon area
Introduce the Objective aperture and redo the Correct astigmatism by CTF step once.
Configuration of the multihole / multishot acquisition