Setup an automated data collection at Glacios

Beam microprobe and nanoprobe calibrations

Prepare the Cross Grating grid for calibration

Microprobe

Focus the beam and center it (with the trackball).

  • Go to microprobe (TEM)
  • Push the Eucentric Focus knob on the console.
  • In the Direct Alignments window:
  1. Adjust the Beam tilt pp X
  2. Adjust the Beam tilt pp Y
  3. Adjust the Beam shift
  4. Adjust the Rotation center
Nanoprobe

Focus the beam and center it (with the trackball).

  • Go to nanoprobe (TEM)
  • Push the Eucentric Focus knob on the console.
  • Check the Condenser 2 centering and adjust if necessary.
  • In the Direct Alignments window:
  1. Adjust the Beam tilt pp X
  2. Adjust the Beam tilt pp Y
  3. Adjust the Beam shift
  4. Adjust the Rotation center

After all calibrations save the settings

Correct astigmatism

  • Verify that the Turbo Auto Off Vacuum option is checked (TEM)
  • Go to x45k Imaging status
  • In the Camera and script window (SEM) open the Setup
  • In Record Tab, set 2 sec exposure, processing unnormalized, align frames, save frames and the objective diaphragm out
  • If the signal is too weak, increase the time, concentrate the beam or change the carbon area
  • Correct astigmatism by CTF
  • Coma free align by CTF
  • Coma vs image shift

Introduce the Objective aperture and redo the Correct astigmatism by CTF step once.


Configuration of the multihole / multishot acquisition