glacios_interface
This is an old revision of the document!
Table of Contents
The Glacios microscope interfaces
The microscope console
Location: right in front of you on the desk
Role: it provides direct control of some of the most important microscope parts:
- Magnification
- Electron dose: spot size (C1 lens) and beam spread (C2 lens)
- Focus (objective lens current)
- Stage position (with the joystick and Z buttons)
- Beam position (with the trackball)
- and many other actions through the programmable switches (L/R1-2-3) and Multifunction X/Y knobs.
Gatan DigitalMicrograph (DM)
Location: On the left screen
Role: It controls the K2 direct electron detector.
- The beginner user will not use it.
SerialRM User Interface (SEMUI)
Location: On the left screen
Role: It provides camera/microscope commands to allow the user to screen its grids and setup an automated data collection.
- ==The beginner user will use it to:|
- record the Atlases.
- select squares and prepare the square maps at eucentric height.
- select targets (holes).
- screen targets (holes).
* ++The expert user will use it to:
- perform all the actions described above.
- perform beam corrections (astigmatism, coma-free).
- perform calibrations.
- setup the dose rate, total dose and dose fractionning.
- setup the time and imaging modes for the different beam modes.
- center the different beam modes all together on the camera axis.
glacios_interface.1644585606.txt.gz · Last modified: (external edit)