Line 1: Line 1:
-Setup an automated data collection at Glacios+====== Setup an automated data collection at Glacios ======
  
-Beam microprobe and nanoprobe calibrations 
  
-Introduce the Cross Grating Grid in the column. +==== Beam microprobe and nanoprobe calibrations ====
-Go to x34 Imaging status. +
-Select 2 positions: a carbon square and an empty square. +
-Introduce the Condenser 2 of 30 um +
-Go to x280 Imaging status. +
-In DigitalMicrograph click View and place the mouse arrow on a feature. +
-Use Alpha Toogle 20° in Stage² window (TEM) to rock the grid. +
-Correct with Z axis knobs on the console. +
-Repeat a second time. +
-Go to x45k Imaging status +
-In Low Dose Control window (SEM) click on Go to: Record. +
-Push the Eucentric Focus knob on the console.+
  
 +++++Prepare the Cross Grating grid for calibration|
 +  * Introduce the Cross Grating Grid in the column.
 +  * Go to x34 Imaging status.
 +  * Select 2 positions: a carbon square and an empty square.
 +  * Introduce the Condenser 2 of 30 um
 +  * Go to x280 Imaging status.
 +  * In DigitalMicrograph click View and place the mouse arrow on a feature.
 +  * Use Alpha Toogle 20° in Stage² window (TEM) to rock the grid.
 +  * Correct with **Z axis knobs** on the console.
 +  * Repeat a second time.
 +  * Go to x45k Imaging status
 +  * In Low Dose Control window (SEM) click on Go to: Record.
 +  * Push the **Eucentric Focus** knob on the console.
 +  * Inset the screen (FC).
 +  * Push the **Wobbler** knob on the console.
 +  * Adjust the Z height on the screen  with the **Z axis knobs** on the console.
 +++++
  
 +== Microprobe ==
  
 +Focus the beam and center it (with the trackball).
 +
 +  * Go to microprobe (TEM)
 +  * Push the **Eucentric Focus** knob on the console.
 +  * In the **Direct Alignments** window: 
 +  - Adjust the ++Beam tilt pp X|<div>
 +  * Click on **Beam tilt pp X** 
 +  * Adjust with **Multifunction X/Y** knobs
 +  * Click **Done** at the bottom of the window.
 +</div>++
 +  - Adjust the ++Beam tilt pp Y|<div>
 +  * Click on **Beam tilt pp Y**.
 +  * Adjust with **Multifunction X/Y** knobs.
 +  * Click **Done** at the bottom of the window.
 +</div>++
 +  - Adjust the ++Beam shift|<div> 
 +  * Click on **Beam shift**.
 +  * Adjust with **Multifunction X/Y** knobs.
 +  * Click **Done** at the bottom of the window.
 +</div>++
 +  - Adjust the ++Rotation center|<div>
 +  * Click on **Rotation center**.
 +  * Adjust with **Multifunction X/Y** knobs.
 +  * Click **Done** at the bottom of the window.
 +</div>++
 +
 +== Nanoprobe ==
 +
 +Focus the beam and center it (with the trackball).
 +
 +  * Go to nanoprobe (TEM)
 +  * Push the **Eucentric Focus** knob on the console.
 +  * Check the Condenser 2 centering and adjust if necessary.
 +  * In the **Direct Alignments** window: 
 +  - Adjust the ++Beam tilt pp X|<div>
 +  * Click on **Beam tilt pp X** 
 +  * Adjust with **Multifunction X/Y** knobs
 +  * Click **Done** at the bottom of the window
 +</div>++
 +  - Adjust the ++Beam tilt pp Y|<div>
 +  * Click on **Beam tilt pp Y**
 +  * Adjust with **Multifunction X/Y** knobs.
 +  * Click **Done** at the bottom of the window
 +</div>++
 +  - Adjust the ++Beam shift|<div> 
 +  * Click on **Beam shift**
 +  * Adjust with **Multifunction X/Y** knobs
 +  * Click **Done** at the bottom of the window
 +</div>++
 +  - Adjust the ++Rotation center|<div>
 +  * Click on **Rotation center**
 +  * Adjust with **Multifunction X/Y** knobs
 +  * Click **Done** at the bottom of the window
 +</div>++
 +
 +++++After all calibrations save the settings|
 +  * Menu: //Settings > Save//
 +  * In **Low Dose Control** window check **Continuous update**
 +  * Click on Update it
 +  * Uncheck **Continuous update**
 +  * Menu: //Settings > Save//
 +++++
 +
 +==== Correct astigmatism ====
 +
 +  * Verify that the **Turbo Auto Off** Vacuum option is checked (TEM)
 +  * Go to x45k Imaging status
 +  * In the Camera and script window (SEM) open the **Setup**
 +  * In Record Tab, set 2 sec exposure, processing unnormalized, align frames, save frames and the objective diaphragm out
 +  * If the signal is too weak, increase the time, concentrate the beam or change the carbon area
 +  * ++Correct astigmatism by CTF|<div>
 +  * Menu: Calibration > Focus & Tuning > Correct astigmatism by CTF
 +  * Open FFT window with Shift+F
 +  * Repeat several times until the corrected parameters have converged
 +</div>++
 +  * ++Coma free align by CTF|<div>
 +  * Menu: Calibration > Focus & Tuning > Coma free align by CTF
 +  * Open FFT window with Shift+F
 +</div>++
 +  * ++Coma vs image shift|<div>
 +  * Menu: Calibration > Focus & Tuning > Coma vs image shift
 +  * Open FFT window with Shift+F
 +  * If the signal is too weak, increase the time, set binning, concentrate the beam or change the carbon area
 +  * parameters for R 1.2/1.3 grids: 5 μm
 +  * parameters for R 2/2 grids: 6 μm
 +</div>++
 +
 +Introduce the Objective aperture and redo the Correct astigmatism by CTF step once.
 +
 +----
 +
 +==== Configuration of the multihole / multishot acquisition ====
 +
 +Do this configuration just before starting the acquisition on one of the squares selected for the whole acquisition after correction of the image shift in x280 and LD mode imaging states.
 +
 +  * On the square map, mark the positions of the 4 holes at the corners of a 9 holes square.
 +  * Go to the center hole with **Go to marker** and properly center on the hole with **Stage shift**.
 +  * Press View in LD mode (ensure that defocus is below -20)
 +  * Open Menu: //Navigation > Montaging & grids > Set multi//
 +  * Change the parameters of the multishot for calibration
 +  * Click on the central hole for positioning
 +  * Adjust the number of shots per hole (in general 3)
 +  * Adjust the distance (72 in general for R 2/2 grids)
 +  * Menu: //Navigation > Montaging & grids > Find hole in regul// to configure
 +  * Select regular pattern 3 x 3
 +  * ++Calibrate after ensuring proper centering on the center hole|<div>
 +  * Move by Image shift to the first corner
 +  * Center properly in the hole using zoom if necessary
 +  * Save the position by clicking on **Save image shift**
 +  * Proceed to the next corner and repeat
 +</div>++
 +  * Save the settings in //Menu > Settings > Save//
 +
 +== Beam tuning ==
 +
 +  * Ensure that LD modes View and Record are perfectly aligned
 +  * Calibrate the proper centering in Record/View/Focus/Trial modes
 + 
 +  * Precisely center Record beam with Direct Alignments/Beam shift
 +  * Check **Continuous update**
 +  * Press View and recenter if necessary with set add beam shift
 +  * ++Then cycle several times between the modes until the beam stop moving|<div>
 +  * Press **Record** and correct Direct Alignments/Beam shift
 +  * Press **Focus** and set add beam shift
 +  * Press **Trial** and set add beam shift
 +  * Repeat
 +</div>++
 +  * Uncheck **Continuous update**
  