 ==== Beam microprobe and nanoprobe calibrations ==== ==== Beam microprobe and nanoprobe calibrations ====
  
 +++++Prepare the Cross Grating grid for calibration|
   * Introduce the Cross Grating Grid in the column.   * Introduce the Cross Grating Grid in the column.
   * Go to x34 Imaging status.   * Go to x34 Imaging status.
Line 20: Line 20:
   * Push the **Wobbler** knob on the console.   * Push the **Wobbler** knob on the console.
   * Adjust the Z height on the screen  with the **Z axis knobs** on the console.   * Adjust the Z height on the screen  with the **Z axis knobs** on the console.
 +++++
  
-Microprobe+== Microprobe ==
  
 Focus the beam and center it (with the trackball). Focus the beam and center it (with the trackball).
Line 49: Line 50:
 </div>++ </div>++
  
-Nanoprobe+== Nanoprobe ==
  
 Focus the beam and center it (with the trackball). Focus the beam and center it (with the trackball).
Line 82: Line 83:
   * In **Low Dose Control** window check **Continuous update**   * In **Low Dose Control** window check **Continuous update**
   * Click on Update it   * Click on Update it
-  * Unchek **Continuous update**+  * Uncheck **Continuous update** 
 +  * Menu: //Settings > Save// 
 +++++ 
 + 
 +==== Correct astigmatism ==== 
 + 
 +  * Verify that the **Turbo Auto Off** Vacuum option is checked (TEM) 
 +  * Go to x45k Imaging status 
 +  * In the Camera and script window (SEM) open the **Setup** 
 +  * In Record Tab, set 2 sec exposure, processing unnormalized, align frames, save frames and the objective diaphragm out 
 +  * If the signal is too weak, increase the time, concentrate the beam or change the carbon area 
 +  * ++Correct astigmatism by CTF|<div> 
 +  * Menu: Calibration > Focus & Tuning > Correct astigmatism by CTF 
 +  * Open FFT window with Shift+F 
 +  * Repeat several times until the corrected parameters have converged 
 +</div>++ 
 +  * ++Coma free align by CTF|<div> 
 +  * Menu: Calibration > Focus & Tuning > Coma free align by CTF 
 +  * Open FFT window with Shift+F 
 +</div>++ 
 +  * ++Coma vs image shift|<div> 
 +  * Menu: Calibration > Focus & Tuning > Coma vs image shift 
 +  * Open FFT window with Shift+F 
 +  * If the signal is too weak, increase the time, set binning, concentrate the beam or change the carbon area 
 +  * parameters for R 1.2/1.3 grids: 5 μm 
 +  * parameters for R 2/2 grids: 6 μm 
 +</div>++ 
 + 
 +Introduce the Objective aperture and redo the Correct astigmatism by CTF step once. 
 + 
 +---- 
 + 
 +==== Configuration of the multihole / multishot acquisition ==== 
 + 
 +Do this configuration just before starting the acquisition on one of the squares selected for the whole acquisition after correction of the image shift in x280 and LD mode imaging states. 
 + 
 +  * On the square map, mark the positions of the 4 holes at the corners of a 9 holes square. 
 +  * Go to the center hole with **Go to marker** and properly center on the hole with **Stage shift**. 
 +  * Press View in LD mode (ensure that defocus is below -20) 
 +  * Open Menu: //Navigation > Montaging & grids > Set multi// 
 +  * Change the parameters of the multishot for calibration 
 +  * Click on the central hole for positioning 
 +  * Adjust the number of shots per hole (in general 3) 
 +  * Adjust the distance (72 in general for R 2/2 grids) 
 +  * Menu: //Navigation > Montaging & grids > Find hole in regul// to configure 
 +  * Select regular pattern 3 x 3 
 +  * ++Calibrate after ensuring proper centering on the center hole|<div> 
 +  * Move by Image shift to the first corner 
 +  * Center properly in the hole using zoom if necessary 
 +  * Save the position by clicking on **Save image shift** 
 +  * Proceed to the next corner and repeat 
 +</div>++ 
 +  * Save the settings in //Menu > Settings > Save// 
 + 
 +== Beam tuning == 
 + 
 +  * Ensure that LD modes View and Record are perfectly aligned 
 +  * Calibrate the proper centering in Record/View/Focus/Trial modes 
 +  
 +  * Precisely center Record beam with Direct Alignments/Beam shift 
 +  * Check **Continuous update** 
 +  * Press View and recenter if necessary with set add beam shift 
 +  * ++Then cycle several times between the modes until the beam stop moving|<div> 
 +  * Press **Record** and correct Direct Alignments/Beam shift 
 +  * Press **Focus** and set add beam shift 
 +  * Press **Trial** and set add beam shift 
 +  * Repeat 
 +</div>++ 
 +  * Uncheck **Continuous update** 