-==== 1.Make sure Microscope is in good shape (ppt associated)==== 
-  - a.Make sure HT and FEG are on 
- * Under « setup » tab,  « high tension » and « operate » are highlighted yellow 
-b. Make sure that column valves are closed and turbo pump is on 
-i. Under « setup » tab,  « col. Valves closed » and « Turbo on » are highlighted yellow 
-c. Make sure that EM column vaccum is stable (IGP1 -6) 
-d. Make sure that Stage neutral 
-i. Under « Search » tab 
-1. click « X,Y » button  to zero out the X, Y to center the stage 
-2. If necessary Click « AB » button  to zero out   coordinates of the stage 
-e. Make sure cold trap already with LN2, refill every 3-4 hours 
  
-2. Load grid in holder 
-a. Place holder in stand with spring holder up 
-b. Lift (open) spring holder using metal pin 
-c. Center specimen grid in holder socket 
-d. d. Close spring holder using metal pin 
  
  
Line 52: Line 36:
 </div>++ </div>++
   - ++Load platform or user settings|<div>   - ++Load platform or user settings|<div>
-  * Menu: //Settings > Open > SerialEM_Settings_SPA//+  * Menu: //Settings > Open > Platform_Settings_SPA//
 </div>++  </div>++ 
  